Open Access

Table 5

Comparison of sensitivity and specificity of Toxo-ISAGA IgM® and ELISA techniques (Architect/Alinity®, Vidas Toxo IgM®, Platelia Toxo IgM®) on 1,206 neonate sera.

Assay (number of sera) Sensitivity % [95% CI] Specificity % [95% CI] p value
Toxo-ISAGA IgM® (n = 1206) 63.1% [59.5–66.5] 98.4% [96.8–99.2] Reference
Architect or Alinity® (n = 285) 48.6% [41.8–55.3] 100% [92.2–100] <0.0001
Vidas Toxo IgM® (n = 177) 17.5% [10.0–28.6] 100% [96.7–100] <0.0001
Platelia Toxo IgM® (n = 344) 64.6% [57.6–71.0] 95.4% [90.8–97.7] 0.2752