Open Access

Table 1

Results of recombinant antigen-based immunochromatography tests (NIE, SsIR, and NIE-SsIR ICT kits) for detection of anti-Strongyloides IgG antibody in human serum samples.

Conditions represented by sera No. of sera tested NIE ICT kit
SsIR ICT kit
NIE-SsIR ICT kit
No. positive2 No. positive2 No. positive2
Group I
Healthy controls 40 0 0 0
Group II
Proven strongyloidiasis 100 81 89 95
Group III
Giardiasis 7 1 2 2
Amoebiasis 10 1 3 0
Blastocystosis 10 2 4 3
Opisthorchiasis 10 5 2 0
Fascioliasis 10 3 1 3
Paragonimiasis 10 2 1 0
Taeniasis 10 2 3 1
Cysticercosis 10 3 2 2
Sparganosis 7 3 2 3
Hookworm infections 10 4 4 1
Ascariasis 10 0 0 0
Trichuriasis 10 2 0 0
Trichinellosis 10 0 2 1
Angiostrongyliasis 10 1 3 2
Gnathostomiasis 10 0 2 0
Capillariasis 10 2 1 1
Diagnostic values [95% CI1]
Sensitivity (%) 81.0 [71.9–88.2] 89.0 [81.2–94.4] 95.0 [88.7–98.4]
Specificity (%) 84.0 [78.1–88.9] 83.5 [77.5–88.4] 90.2 [85.1–94.0]
Positive likelihood ratio 5.07 [3.62–7.10] 5.40 [3.90–7.46] 9.70 [6.31–14.90]
Negative likelihood ratio 0.23 [0.15–0.34] 0.13 [0.08–0.23] 0.06 [0.02–0.13]
1

CI, confidence interval.

2

The results were correlated between interpretations by the naked eye and the in-house strip reader.

Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.

Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.

Initial download of the metrics may take a while.