Performance comparison between LAMP and microscopy or real-time PCR. Sensitivity, specificity, positive predictive value and negative predictive value were calculated based on 299 samples, excluding the 11 invalid results from LAMP. The number of samples used for microscopy are 79 positive (83–4 invalid excluded) and 220 negative (227–7 invalid excluded). The number of samples used for PCR are 85 positive (89–4 invalid excluded) and 214 negative (221–7 invalid excluded).

Microscopy Real-time PCR


+ total + total
LAMP + 79 14 93 85 4 89
0 206 206 0 210 210
79 220 299 85 214 299
Versus microscopy Versus real-time PCR


 LAMP performance % 95% confidence interval (%) % 95% confidence interval (%)
Sensitivity 100 95.55 to 100 100 95.75 to 100
Specificity 93.64 89.55 to 96.48 98.13 95.28 to 99.49
Positive predictive value 84.95 77.27 to 90.35 95.51 88.95 to 98.25
Negative predictive value 100 n.a. 100 n.a.
Positive likelihood ratio 15.7 n.a. 53.5 n.a.